LCR40 - Atlas LCR Passive Component Meter Information:
The Atlas LCR40 automatically identifies and analyses almost any passive component (Inductor, Capacitor or Resistor). It measures its parameters using automatically selected test frequencies (DC, 1kHz, 15kHz and 200kHz).
Inductance from 1uH to 2H (min resolution is 0.8uH), Capacitance from 1pF to 10000uF (min resolution is 0.5pF) and Resistance from 1ohm to 2Mohm.
Includes:
Resistance: 1 Ohm to 2 MegOhms, resolution typically 0.6 Ohms, accuracy typically ±1.0% ±1.2 Ohms
Inductance: 1uH to 2H, resolution typically 0.8uH, accuracy typically ±1.5% ±1.6uH
Capacitance: 0.5pF to 10000uF, resolution typical 0.5pF, accuracy typically ±1.5% ±1.0pF
Open circuit test voltage: ±1.05V
Short circuit test current: ±3.25mA
Test frequencies: DC, 1kHz, 14.925kHz and 200kHz, ±1.5%
Sine purity: -60dB 3rd harmonic typical
Operating temperature range: 15°C to 35°C (60°F to 95°F)
Battery type: GP23A or V23A or MN21 (12V 55mAH type)
Battery life: Typically ~1600 operations at <1 minute per operation
Display type: Alphanumeric LCD (not backlit)
Please avoid charge/voltage on the probes.
Accuracies shown for 100uH-100mH, 200pF-500nF and 10 Ohms to 1MOhms.
Inductance from 1uH to 2H (min resolution is 0.8uH), Capacitance from 1pF to 10000uF (min resolution is 0.5pF) and Resistance from 1ohm to 2Mohm.
Includes:
- LCR40 Passive Component Analyser.
- 2mm plugs and sockets and removeable hook probes.
- Comprehensive illustrated user guide.
- GP23 Alkaline battery.
Technical Specifications
Resistance: 1 Ohm to 2 MegOhms, resolution typically 0.6 Ohms, accuracy typically ±1.0% ±1.2 Ohms
Inductance: 1uH to 2H, resolution typically 0.8uH, accuracy typically ±1.5% ±1.6uH
Capacitance: 0.5pF to 10000uF, resolution typical 0.5pF, accuracy typically ±1.5% ±1.0pF
Open circuit test voltage: ±1.05V
Short circuit test current: ±3.25mA
Test frequencies: DC, 1kHz, 14.925kHz and 200kHz, ±1.5%
Sine purity: -60dB 3rd harmonic typical
Operating temperature range: 15°C to 35°C (60°F to 95°F)
Battery type: GP23A or V23A or MN21 (12V 55mAH type)
Battery life: Typically ~1600 operations at <1 minute per operation
Display type: Alphanumeric LCD (not backlit)
Please avoid charge/voltage on the probes.
Accuracies shown for 100uH-100mH, 200pF-500nF and 10 Ohms to 1MOhms.